Application note · Food & Beverage

Using UV-Vis for Colour and Routine Quantitation

A method-first guide to fixed-wavelength colour tests and routine absorbance work in food QC, including when a colorimeter is enough and when a spectrophotometer earns its place.

Yoke Application Team · Reviewed by Yoke Application Team · Updated October 22, 2026 · 3 min read

The short answer

  • Colour is a perception, so a number that stands for it has to name the scale: fixed-wavelength transmittance, platinum-cobalt, or tristimulus values.
  • A colorimeter is built for one job at production frequency; a spectrophotometer adds the ability to find a wavelength and run a second method.
  • Colour standards fade quietly — give them their own check and replacement interval.

Colour and concentration are two different questions asked of the same platform. Colour asks whether the sample matches a reference; concentration asks how much of a substance is present. Both can be answered by absorbance, and confusing the two is where routine food QC goes wrong.

Colour is not a wavelength

Colour is a perception, and a number that stands for it has to be defined: transmittance at a fixed wavelength, a platinum-cobalt value, or a set of tristimulus values. Each is valid and each answers a different question. A specification that says “colour within limit” without naming the scale is not measurable.

Fixed-wavelength quantitation

For a single known analyte, a fixed wavelength and a calibration curve are enough. The discipline sits in the details: blank measured in the same cell, wavelength verified, curve bracketing the samples, and every dilution recorded with the result.

Two tasks, one platform

Task Measurement approach Common error
Colour against a reference Transmittance at a fixed wavelength, or a colour scale Comparing values produced at a different path length
Colour of a turbid sample Filter or centrifuge, then measure Reading the sample with particles in suspension
Single analyte, routine Calibration curve at a fixed wavelength Sample outside the calibrated range
Scanning for a maximum Spectral scan before fixing the wavelength Choosing a wavelength from a published method without checking it on this matrix

Colorimeter or spectrophotometer

A colorimeter is built for one job and does it quickly with a fixed optical geometry — appropriate when the measurement is a colour scale at production frequency. A spectrophotometer covers the same measurement and adds the ability to locate a wavelength, scan a spectrum and run a second method. Choose by the number of methods, not by the number of samples.

Calibration for colour

Colour standards drift and fade. A reference material that has aged quietly shifts every result that depends on it, so the standards need their own check and their own replacement interval. Write both down, and record the standard’s identifier with the result.

Cell handling

In colour work the cell is part of the method. Path length, cleanliness, filling height and orientation all change the reading. Keep one cell per product where possible, check it against a blank at the start of the shift, and handle it by the frosted faces only.

Where colour results go wrong in practice

Four failures recur. A sample read without filtration, so particles add apparent absorbance. A cell used in a different orientation from the one it was filled in. A reference standard that has aged past its own validity. And a result compared with a value produced on a different instrument type, where the optical geometry differs enough to change the number.

All four are cheap to prevent and hard to detect afterwards, because each produces a plausible value rather than an error message.

The brief

Send the colour scale required, the sample form and path length, the number of methods the laboratory runs, the sample frequency and the record format. Those five decide the platform.

Instruments used in this note.

The configuration described above, with the full specification of each model.

1000 Series Basic Spectrophotometer

5 models

1000 Series Basic Spectrophotometer

Five basic single-beam spectrophotometers from V1000 to UV1100N, with a 1200 lines/mm grating, for routine visible and UV-Vis measurement.

Optical System
Single Beam, Grating 1200 lines/mm
Wavelength Repeatability
1nm
Photometric Range
0-200%T, -0.3-3.0A
View specifications

Questions about this method.

Can a spectrophotometer replace a colorimeter for colour grading?

It can produce the same measurement and more, at a lower throughput per sample and a higher cost. Where the only task is a colour scale at production frequency, a dedicated colorimeter is the more practical instrument.

Why does the same sample give different colour values on two instruments?

Different optical geometry, different cell path, or a different scale. Colour numbers are only comparable when the instrument type, the path length and the scale are the same — which is why the specification has to name all three.

Does turbidity affect a colour reading?

Yes. Particles scatter light and add apparent absorbance, so a turbid sample reads darker than it is. Filter or centrifuge before measurement, and state which in the method.

More notes on this application.

All notes

Let’s talk about your application

Check the configuration against your method.

Send the sample, the concentration range and the reference method. We will confirm the instrument and the accessories.