IT Series
IT-600U Double Beam Spectrophotometer
1-inch touchscreen and windows graphic interface.
Documentation
Datasheets and manuals are issued per model and revision. Ask for the documents that match the configuration you are considering.
See linked documents| Optical System | Double Beam, Grating 1200 lines/mm |
|---|---|
| Wavelength Range | 190-1100 nm |
| Spectral Bandwidth | 1.8nm |
| Wavelength Accuracy | ±0.1nm@656.1nm, ±0.3nm@all |
| Wavelength Repeatability | <=0.1nm |
| Photometric Accuracy | 0.2%T(0~100%T), ±0.002A(0-0.5A), ±0.004A(0.5-1A) |
| Photometric Repeatability | ≤0.15%T (0-100%T), 0.001A(0-0.5A), 0.002A(0.5-1A) |
| Photometric Range | 0-200%T, -0.3~3A, 0-9999C (0-9999F) |
| Stray Light | <=0.03%T@220nm, 360nm |
| Stability | ±0.0003A/h @500 nm |
| Baseline Flatness | ±0.001A |
| Noise | 0.0005A@500nm |
| Work Mode | T, A, C, E |
| Scanning Speed | Hi, Med, Low (Max. 3000nm/min) |
| Wavelength Setting | Auto |
| Display | 10.1-inch Touch Screen |
| Light Source | Deuterium &Tungsten lamp |
| Detector | Silicon Photodiode |
| Cuvette Holder | 10mm single hole cell holder |
| Output | USB drive, USB host, RS232 |
| Power | AC 220V/50Hz or AC 110V/60Hz |
Send your sample type and method, and we will confirm the configuration and current lead time.
Ask about this modelOverview
What this instrument is designed to do, in plain terms.
- 10.1-inch touchscreen with a Windows graphical interface.
- Functions include Photometric measurement, Quantitative measurement, Kinetics, Spectrum scan, DNA/Protein test, multi-wavelength test, etc.
- In-house massive memory is capable of saving up to 1024M for test data & working curves.
- Supports USB storage. The USB port can be used for data transfer and exporting to Excel for further processing, analysis and storage.
- Socket-type deuterium and tungsten lamps support lamp switching without optical realignment, and the lamps are easy to replace.
- Extensive accessories are optional, such as film holder, integrating sphere, peltier/sipper system, 21 CFR compliant software, built-in thermal printer, PC software etc.
Accessaries
Confirm these items before the final quotation. Accessaries are listed only where the manufacturer offers them.
Standard Accessaries
- 10mm glass cuvette x 4
- 10mm quartz cuvette x 2
- Power cord
- User's Manual
Optional Accessaries
- film holder
- integrating sphere
- peltier/sipper system
- 21 CFR compliant software
- built-in thermal printer
- PC software
Specifications can depend on the selected configuration. Confirm the current datasheet, accessories and software version before purchase.
Let’s talk about your application
Find the right instrument for your application.
Share your sample, measurement requirements and preferred configuration.